A-1000 Performance Specification

Performance
Automatic TOC Mode--
Operating Range 0.05 to 9999 ppb as carbon
Repeatability:
±0.05 ppb < 5 ppb TOC
±1% > 5 ppb TOC
Minimum input Resistivity:
5.0MW -cm for all waters
1.0MW -cm for neutral waters
0.2MW -cm for water with CO2 as sole
conductive species
Ambient Operating Temperature:
S10 Sensor: 5°C to 40°C (41°F
to 104°F)
S20/S20P Sensor: 5°C to 35°C
(41°F to 95°F)
Sample Water Temperature: 0°C to 100°C (32°F to 212°F)
Inlet Pressure: 100 psig maximum(690 kPa)
Display Resolution:
0.00 to 19.99 ppb
20.0 to 1999.9 ppb
200 to 9999 ppb
Purge ( Resistivity / Conductivity ) Mode ---
Operating Range:
Resistivity: 0.01 to 18.2 MW-cm
Conductivity: 0.05 to 100 microsiemens/cm
Temperature Compensation: To 25°C over entire range of 0°C ~100°C (based on a
NaC1 model)
Display Resolution: Three significant figures as resistivity
Four significatant figures as conductivity
Accuracy: ±3% over 0°C ~100°C range
Physical
A-1000 Network Capacities --
Type: RS-485
Maximum Sensors: 8
Maximum C80 Controllers(any configuration): 8
Maximum Network Length: 3000 feet (1 km)
Network Cabling: Shielded Twin-axial, Twist-Lock BNC
C80 Controller

The Model C80 Controller
provides multiple Sensor display, individual instrument setup, system alarm
indication and printer control functions for up to 8 Sensors via an RS485
Network.
C80 Controller P/N FG 1000301[ 120 VAC, 60 Hz]
P/N FG 1000301[ 230 VAC, 50 Hz]
Displays ---
Main: 4-line x 16-character Super-Twist LCD
Channel: 8 Bi-color(red/green)LCDs
Physical ---
Operating Temperature: 0°C to 40°C (32°F to 104°F)
Case: ABS front panel with Lexan overlay
Size: 120mm L x 81mm W x 48mm D
(4.7-inch L x 3.2-inch W x 1.9-inch D)
2.5-inch total depth behind panel (allow 4-5-inch to avoid crimping)
Weight: 0.75 kg(1.5 lbs)
Mounting: Panel
Connections ---
Communications: Shielded Twin-axial, Twist-Lock BNC
Power: 9 VDC Wall-Mount Power Pack
(120 VAC, 60 Hz standard, 230 VAC, 50 Hz optional)
S10 Sensor P/N FG10000401
The Model
S10 Sensor is the basic analysis component in the A-1000 TOC Analytical
System.
Display --
Main: 1-line x 16-character Super-Twist LCD
Backlighting: Yellow LED
Character Height: 0.163-inch
Physical --
Operating Temperature: 0°C to 35°C (32°F to 95°F) Size: 330mm L x 172 mm W x
112 mm D
(13.0-inch L x 6.8-inch W x 4.4-inch D )
Weight: 6.5 kg. (12.75 lbs.)
Power: 85 to 264 VAC, 50 to 60 Hz
I/O Connections --
Serial Interfaces: RS485 opto-isolated Network
Rs232 Data Acquisition
Rs232 Diagnostics
Digital I/O: Two opto-isolated inputs
Two opto-isolated outputs
Analog: Opto-isolated 4-20 mA output
Non-isolated 12 VDC output @ 1/2 Amp max.
S20 Sensor P/N FG 1000501
The Model
S20 Sensor combines the standard Model S10 Sensor, a C80 Controller, a handle
and feet to furnish a portable TOC monitoring instrument.
Display --
Main: 4-line x 16-character Super -Twist LCD
Backlighting(adjustable): Yellow LCD
Character Height: 0.163-inch
Physical --
Operating Temperature: 0°C to 35°C (32°F to 95°F)
Size: 330mm L x 172mm W x 112mm D
(13.0-inch L x 6.8-inch W x 4.4-inch D)
Weight: 7 kg. (14lbs.)
Power: 85 to 264 VAC, 50 to 60 Hz
Connections --
Serial Interfaces: RS485 opto-isolated Network
Rs232 Data Acquisition
RS232 Printer
RS232 Diagnostics
Digital I/O: Two opto-isolated inputs
Two opto-isolated outputs
Analog: Opto-isolated 4-20mA output
S20P Sensor P/N FG 1000501

The Model S20P Sensor
combines a C80 Controller, handle, feet ,the standard Model S10 Sensor and a
40-column thermal printer to provide a complete and integrated portable TOC
analysis system.
Display --
Main: 4-line x 16-character Supter-Twist LCD
Backlighting(adjustable): Yellow LCD
Character Height: 0.163-inch
Physical --
Operating Temperature: 0°C to 35°C (32°F to 95°F)
Size: 330mm L x 172mm W x 112mm D
(13.0-inch L x 6.8-inch W x 4.4-inch D)
Weight: 8.25 kg. (16.5lbs.)
Power: 85 to 264 VAC, 50 to 60 Hz
Connections --
Serial Interfaces: RS485 opto-isolated Network
Rs232 Data Acquisition
RS232 Diagnostics
Digital I/O: Two opto-isolated inputs
Two opto-isolated outputs
Analog: Opto-isolated 4-20mA output
A1000XP TOC Analyzer Specifications (For Semiconductor)
( for
Semiconductor)

1. No reagent, No Gas.
2. Operation Range: 0.020 ppb ~ 1,999 ppb.
3. Accuracy :
±0.05 ppb, 0.02 ppb ~ 2.000 ppb
±0.1 ppb, 2.000 ppb ~ 9.999 ppb
±2% , Greater than 10.00 ppb
4. Reapeatability: ±0.05 ppb.
5. Resolution:
0.001 ppb, 0.020ppb ~ 9.999
ppb.
0.01 ppb, 10.00ppb ~ 19.99
ppb
0.1 ppb, 20.0 ppb ~ 199.9 ppb
1 ppb, Greater than 200 ppb.
6. Display: T.O.C., Resistivity/ Conductivity, Temperature, Trend, Profile.
7. Power Supply: 90 ~ 230 VAC.
Anatel A-643P Specification
TOC Mode--
Operating Range: 1 to 1,999 ppb as carbon
Resistivity:
0.01 to 18.2 MΩ-cm
Conductivity:
0.05 to 150μS/cm(@25°C)
Display Resolution: 1 ppb
Sample
Water Temperature: 5°C to 95°C (41℉
to 203 ℉ )
Conductivity(Purge)Mode--
Conductivity Mode: Temperature Compensated to 25°C
or Uncompensated
Conductivity
Range: 0.05 to 150μS/cm(@25°C)
Display
Resolution: 0.01μS/cm
over full range
Conductivity Accuracy‡: 2% over Full range
(Uncompensated)
● †Repeatability
of reading on an individual instrument
● ‡Repeatability of
reading from instrument-to-instrument
Calibration & Validation--
TOC—
Up to Four-Point Calibration: 0 ppb C (blank), 250 ppb C,
500 ppb C, 750 ppb C (all sucrose)
System
Suitability: 500 ppb C (sucrose)
500
ppb C (1-4,benzoquinone) Reagent Blank
Conductivity--
One-Point
Calibration, Cell Constant: 100μS/cm solution
One-Point
Validation of Meter: 61.9KW resistor


A-2000 Performance Specifications
PERFORMANCE Specification
Operating
Range(as total carbon) --
Range
1: 20 ppb to 10,000 ppb
Range
2: 1.0 ppm to 100.0 ppm
Range
3: 100.0 ppm to 1,000 ppm
Accuracy
--
Range
1:±2% or 5ppb(whichever is greater)
Range
2:±2% or 0.5 ppm(Whichever is greater)
Range
3:±2% or 0.5 ppm(Whichever is greater)
Sparge Mode ( TOC ) –
Report Total Organic Carbon ( non-purgeable)
Range 1: MDL ( 40μg/L ) to 10mg/L TOC
Range 2: 1 mg/L to 100 mg/L TOC
Range
3: 100 mg/L to 1000 mg/L TOC
Differential Mode(TOC [TC – TIC])—
Reports
Total Carbon (TC), Total Organic Carbon (TOC) and
Total Inorganic Carbon (TIC)
Range
1: MDL (20μg/L ) to 10 mg/L TC
(use
Sparge Mode if TIC>TOC or TIC > 5 mg/L)
Range
2: 1 mg/L to 100 mg/L TC
(
use Sparge Mode if TIC>TOC or if TIC>10mg/L)
Range
3: 100 mg/L to 1000 mg/L TC
(
use Sparge Mode if TIC>TOC or if TIC > 50mg/L)
Total Carbon Mode (TC ) --
Reports
Total Carbon ( TOC + TIC )
Range
1: MDL ( 20μg/L ) to 10 mg/L TC
Range
2: 1 mg/L to 100 mg/L TC
Range
3: 100 mg/L to 1000 mg/L TC
Total Inorgainc Carbon Mode (TIC)--
Reports
Total Inorganic Carbon (or carbonate)
Range
1:MDL(20μg/L ) to 10 mg/L TIC
Range
2: 1mg/L to 100 mg/L TIC
Range
3: 100 mg/L to 1000 mg/L TIC
Disk Drive—standard
Type:
3.5-inch, 1.44Mb floppy
Printer—standard
Internal: 23-Column Thermal
Sample Introduction-- Optional
On-Line, Manual or Auto Sampler
I/O Connections—
Serial
Interface: RS-232
Anet:
RS-485
External
Printer:Parallel (Epson 9- or 24-pin),
Hewlett-Packard
LaserJet or DeskJet)
Analog: Four opto-isolated 4-20mA Outputs
Digital:
Two opto-isolated inputs
Five
Alarm Relay outputs
Anet Network—
Type:
RS-485
Maximum
Sensors:8
Maximum
Network Length: 3000 feet(1 km)
Network
Cabling:
Shielded Twin-axial, Twist-Look BNC


Taiwan
Hallmark Corporation
Phone: (02) 2659-5006 (6 lines) Fax: (02) 2659-5012
Address: 6FL, NO.50, Lane 258 , JuiKuang Rd. NeiHu, Taipei, Taiwan
E-mail: info@anatel.com.tw
http://www.anatel.com.tw
|